Description

It is a homodyne detector with a low pass filter. An external reference signal is required to operate SU-1. All operating parameters (amplification, time constant and phase) are manually adjustable. Also this detector is used to measure phase shift of the signal.

The principle of operation of SU-1 is based on orthogonality of two sinusoidal functions. When a product of two sinusoidal functions with non-matching frequencies f1 and f2 is integrated over time well above their periods, the result is zero. On the other hand, when f1 = f2, and both functions are in phase, integrated average is equal to half of the product of the functions’ amplitudes.

SU-1 multiplies input signal to reference signal (from built-in generator or external source) and integrates the product over specified time (usually from milliseconds to a few seconds). The result is a DC signal, and the effect of any other signal not matched by frequency to the reference is dampened to almost zero. Same-frequency component that is not in phase with the reference signal is attenuated too (since same-frequency sinusoidal and cosinusoidal waves are orthogonal).


Key features:

Operating frequency up to 250 KHz;
Adjustable sensitivity, time constant and phase;
Compact and EMI (electromagnetic interference) shielded enclosure.

SU-1 package includes:

electronic unit,
power supply,
3 RF-connector cables (1.5 m each),
operating mannual.

Following accessories for SU-1 can be supplied separatly:

Golay Detector GC-1P/T/D;
BPF (band pass filters) for a specified wavelength within 0.1-15 THz range;
LPF (low pass filters) to filter out IR radiation with the following cut off frequencies: 23.4 THz, 23.3 THz, 23.1 THz, 14.3 THz, 10.9 THz, 8.8 THz, 5.5 THz, 4.3 THz, 4 THz, 3.2 THz;
a set of attenuators with 1%, 3%, 10% и 30% transmittence;
TPX and HRFZ-Si lenses.

SU-1 is to be used to amplify high-noise signal produced by the following means of detecting of THz radiation: electro-optical detection of pulsed THz radiation and of detection of THz radiation using the following detectors: Golay detectors, cooled bolometers, pyro-electric THz radiation detectors, semiconductor THz radiation detectors, novel THz radiation detectors.